Bronkhorst FLOW-BUS interface Manuale Utente Pagina 20

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 19
BRONKHORST®
Page20
FLOWBUSinterface9.17.024
4.1.3 RemoteinstallationtoFLOWBUS
Some FLOWBUS modules have to be installed on the bus by means of another FLOWBUSmodule or a (personal)
computer.Thiscanbedoneasfollows:
WithaReadout/Controlmodule,seriesE7000orwithasoftwareprogramonthe PC thisnewmodulecanbe
searched.
The Reado
ut/Controlmoduleor PCprogramwill findthe firstfree nodeaddresscountedfrom address 3 (3 is
firstfreeaddress).
Thisfreeaddressnumberwillbesendtothenewmodule,whichshouldbecomeactiveonthisaddress.
After a reset command, also send by the Readout/Co
ntrol module or PCprogram, the new module will be
operationalontheFLOWBUSsystem.
Makesuretoinstallonlyonenewmoduleatthesametime.
4.1.4 Installtobusonapredefinednodeaddressorreaddressinstrumentonthebus
InsomeapplicationsitisnecessarytoputFLOWBUSdevicesonpredefinedaddressesbecauseoftheapplication
softwareexpectingthisdeviceonthisaddress(e.g.PLCapplications).Thiscanbedoneasfollows:
(Installnewmoduletothebusasdescribedinthepreviousparagraphs.)
Bymeansoftherotaryswitchonth
einstrument(ifpresent).
Bymeansofthemicroswitchontheinstrument.Seeinstructionmanual9.17.023.
By means of the special menu in an E7000 module for digital instruments it is possible to (re) address
instrumentsonthebus.
Gototh
especificmenupartandreadthePNA(PrimaryNodeAddress)number.
Fillintheaddressyouwanttheinstrumenttohaveonthebus.
TheE 7000modulewillcheckifthisaddressisalreadyoccupiedbyanotherinstrument.Ifso,youcouldremove
theotheroneandtryag
ain,orenteranotheraddress.
Themodulewillrestartandwillgetitsnewaddressfromthatmomenton.
BecauseitisnotallowedinaFLOWBUSsystemtohave2devicesonthesamenodeaddress,at
(re)startoftheinstrumenttherewillalsobeacheckperformedifitsnodeaddress(inmemory)is
alreadyoccupiedbyanotherdevice.Ifso,theredLEDwillblinkslowly.Atthatmom
entthemoduleis
notoperationalonthebus.Youfirsthavetoreinstallit.Theeasiestwaytodothisistopressthemicro
pushbuttonontopoftheinstrumentuntilthegreenLEDstartsblinkingfast.Afterreleasingthebutton
theinstrumentwillinstallitselfonth
ebusautomaticallyonafreeaddress.WhenthegreenLEDis
burningcontinuously,theinstrumentisoperational.
Ifyoutakeoneormoremodulesoutofyoursystem,orifoneormoremodulesdonotgetpower,be
carefulinstallingnewmodules,orchangingtheconfiguration.Thissituationmayoccurifyoutake
modulesoutforservice.Ifyoudon’tchangetheconfiguration,youcanreplaceyourse
rviced
instrument/moduleinyoursystemwithoutanyproblem.Mostlynewmoduleswillgetthefirstfree
address,countedfromaddress3(3willbethefirst).Thiscouldbeanaddresswhichshouldnotbe
occupied(e.g.fromamodulewhichhasbeentakenoutforservice).Inthatcaseinstallth
enewmodule
andreaddressitasdescribed,toavoidFLOWBUScommunicationproblems.
Whenconnectinganinstrument/moduletotheFLOWBUS,makesurethattheinstrument/moduleisin
poweroffmode.Afterconnectiontothebusyoucanswitchpoweron.Atpoweruptherewillbea
checktoavoidthesituationoftwomodulesonthesamebusaddress.Thisisimportantfo
rerrorfree
buscommunication.